醫(yī)用電子內(nèi)鏡常見(jiàn)故障的處理

來(lái)源:http://wufangbu88888.cn/ 發(fā)布時(shí)間:2023-06-17 瀏覽量:0

電子內(nèi)鏡按鈕失靈
Electronic endoscope button malfunction
電子內(nèi)鏡操縱部按鈕的表面膠皮很薄,在使用和清洗中容易破損產(chǎn)生針孔而導(dǎo)致按鈕進(jìn)水失靈。從按鈕的外觀上很難發(fā)現(xiàn)細(xì)小的針孔,因此,在內(nèi)鏡清洗之前要進(jìn)行測(cè)漏,以便及時(shí)發(fā)現(xiàn)漏水的部件進(jìn)行更換,防止故障進(jìn)一步擴(kuò)大。
The surface rubber of the electronic endoscope control button is very thin, which is prone to damage and pinholes during use and cleaning, leading to water ingress and malfunction of the button. It is difficult to detect small pinholes from the appearance of the button, therefore, leakage testing should be carried out before endoscopic cleaning to timely detect and replace leaking components and prevent further expansion of the fault.
電子內(nèi)鏡電氣接頭進(jìn)水
Water ingress into the electrical connector of the electronic endoscope
內(nèi)窺鏡設(shè)備維修人員認(rèn)為其電子內(nèi)鏡電氣接頭進(jìn)水的原因是清洗時(shí)忘蓋防水帽或防水帽故障,導(dǎo)致液體進(jìn)入電氣接頭腐蝕插座,造成電子內(nèi)鏡無(wú)法向圖像處理器傳輸信號(hào)。因此,需要定期對(duì)防水帽的密封性進(jìn)行檢查,及時(shí)維修或更換相關(guān)部件。
The maintenance personnel of the endoscope equipment believe that the reason for the water ingress into the electrical connector of the electronic endoscope is due to forgetting to cover the waterproof cap during cleaning or a malfunction of the waterproof cap, which causes liquid to enter the electrical connector and corrode the socket, causing the electronic endoscope to be unable to transmit signals to the image processor. Therefore, it is necessary to regularly inspect the sealing of the waterproof cap and repair or replace relevant components in a timely manner.
內(nèi)窺鏡設(shè)備維修
電子內(nèi)鏡圖像異常
Abnormal electronic endoscope image
電子內(nèi)鏡圖像經(jīng)常出現(xiàn)的問(wèn)題有圖像偏暗、顏色異常、圖像時(shí)有時(shí)無(wú)等,將直接影響醫(yī)生的診斷。圖像偏暗時(shí)檢查冷光源的亮度、內(nèi)鏡導(dǎo)光束的情況、監(jiān)視器亮度及對(duì)比度的設(shè)置,并作相應(yīng)調(diào)整;顏色異常時(shí),需重新做白平衡,并檢查監(jiān)視器亮度及對(duì)比度的設(shè)置;圖像時(shí)有時(shí)無(wú)往往是內(nèi)鏡連接圖像處理器的電纜接觸不良所致,此時(shí),可通過(guò)處理電纜的電氣接口來(lái)排除故障。
The common problems that occur in electronic endoscope images include dim images, abnormal colors, and sometimes absent images, which directly affect the doctor's diagnosis. When the image is dim, check the brightness of the cold light source, the condition of the endoscopic guide beam, the setting of the monitor brightness and contrast, and make corresponding adjustments; When the color is abnormal, it is necessary to redo the white balance and check the brightness and contrast settings of the monitor; Sometimes the lack of images is often caused by poor contact of the cable connecting the endoscope to the image processor. In this case, the fault can be eliminated by processing the electrical interface of the cable.
內(nèi)鏡的日常維護(hù)及檢測(cè)工作十分重要,要做到小心使用,經(jīng)常測(cè)漏,發(fā)現(xiàn)問(wèn)題及時(shí)處理,便可以大限度地減少內(nèi)鏡維修次數(shù),延長(zhǎng)其使用壽命。更多相關(guān)事項(xiàng)就來(lái)我們網(wǎng)站http://wufangbu88888.cn咨詢!
The daily maintenance and testing of endoscopes are very important. It is important to use them carefully, frequently detect leaks, and promptly handle any problems found. This can greatly reduce the number of endoscopic repairs and extend their service life. For more related matters, come to our website http://wufangbu88888.cn consulting service
s
新聞資訊
聯(lián)系我們
  • 公司公司:匠仁醫(yī)療設(shè)備有限公司
  • 匠仁醫(yī)療山東濟(jì)南維修部
  • 聯(lián)系人:樊經(jīng)理
  • 電話:13153199508
  • 地址:山東省濟(jì)南市槐蔭區(qū)美里東路3000號(hào)德邁國(guó)際信息產(chǎn)業(yè)園6號(hào)樓101-2室
  • 匠仁醫(yī)療湖南長(zhǎng)沙維修部
  • 聯(lián)系人:李經(jīng)理
  • 電話:13873135765
  • 地址:湖南省長(zhǎng)沙市雨花區(qū)勞動(dòng)?xùn)|路820號(hào)恒大綠洲14棟2409室
新聞資訊

樊經(jīng)理:13153199508     李經(jīng)理:13873135765

公司地址:山東省濟(jì)南市槐蔭區(qū)美里東路3000號(hào)德邁國(guó)際信息產(chǎn)業(yè)園6號(hào)樓101-2室  湖南省長(zhǎng)沙市雨花區(qū)勞動(dòng)?xùn)|路820號(hào)恒大綠洲14棟2409室

     

魯公網(wǎng)安備 37010402441281號(hào)