醫(yī)用內(nèi)窺鏡系統(tǒng)的組成與故障維修

來(lái)源:http://wufangbu88888.cn/ 發(fā)布時(shí)間:2024-04-16 瀏覽量:0

(1)光源系統(tǒng)常見(jiàn)的故障

(1) Common faults in light source systems

(a)光線(xiàn)發(fā)暗。排除了是光柵遮擋的原因外,基本可以肯定是燈泡接近使用壽命,更換新燈泡則可。氙光燈泡的使用壽命一般設(shè)計(jì)在500小時(shí),如果光源系統(tǒng)有壽命計(jì)時(shí)裝置,這個(gè)問(wèn)題就很容易。

(a) The light is dim. Excluding the possibility of grating obstruction, it is generally certain that the bulb is nearing its lifespan, and replacing it with a new bulb is sufficient. The service life of xenon light bulbs is generally designed to be 500 hours. If the light source system has a lifespan timer device, this problem can be easily solved.

(b)開(kāi)機(jī)燈泡不亮。如果排除了燒掉燈泡這個(gè)原因的話(huà),那么問(wèn)題常常就出在主機(jī)高壓點(diǎn)火電路部分,因?yàn)殚_(kāi)機(jī)的時(shí)候機(jī)器會(huì)產(chǎn)生上萬(wàn)伏特的脈沖高壓,元件會(huì)老化得特別快。注意此時(shí)不能用萬(wàn)用表去檢查脈沖高壓。

(b) The light bulb does not light up when turned on. If the reason for burning out the light bulb is ruled out, the problem often lies in the high-voltage ignition circuit of the host, because when the machine is turned on, it will generate tens of thousands of volts of pulse high voltage, and the components will age very quickly. Be careful not to use a multimeter to check for pulse high voltage at this time.

(2)人工氣腹系統(tǒng)常見(jiàn)的故障

(2) Common malfunctions of artificial pneumoperitoneum systems

(a)壓力顯示不正確。這多是氣體壓力傳感器損壞,如果機(jī)器設(shè)計(jì)有一組多個(gè)壓差壓力傳感器同時(shí)工作的,一般一個(gè)老化或者污染無(wú)法工作的時(shí)候,我們要把這組傳感器全部更換,否則只換其中一個(gè),其他的同組的傳感器壽命也接近臨界了。

(a) The pressure display is incorrect. This is mostly due to damage to gas pressure sensors. If the machine is designed with a group of multiple differential pressure sensors working simultaneously, usually when one sensor is aging or contaminated and cannot work, we need to replace all of this group of sensors. Otherwise, only one of them will be replaced, and the lifespan of other sensors in the same group will also be close to the critical point.

(b)流量不準(zhǔn)確。氣體流量傳感器損壞更換即可。有時(shí)候是通過(guò)壓差來(lái)計(jì)算流量的,這時(shí)候我們就要更換新的壓差傳感器。

20200426035244690

(b) Inaccurate traffic. Replace the damaged gas flow sensor. Sometimes the flow rate is calculated by pressure difference, and at this time we need to replace it with a new pressure difference sensor.

(c)設(shè)備漏氣。很多情況是器械密封圈老化而引起的。但也有設(shè)備機(jī)械減壓閥漏氣和電磁閥損壞的原因,這個(gè)比較容易判斷出并對(duì)其進(jìn)行修復(fù)。

(c) Equipment leaks air. Many cases are caused by aging of instrument sealing rings. But there are also reasons for mechanical pressure reducing valve leakage and electromagnetic valve damage, which are relatively easy to identify and repair.

(3)液體加壓系統(tǒng)常見(jiàn)的故障

(3) Common faults in liquid pressurization systems

(a)蠕動(dòng)泵損壞。

(a) The peristaltic pump is damaged.

(b)液體壓力傳感器損壞。其原理基本與氣體壓力傳感器相同。

(b) The liquid pressure sensor is damaged. Its principle is basically the same as gas pressure sensors.

(c)設(shè)備內(nèi)部電路損壞。

(c) The internal circuit of the equipment is damaged.

(4)電凝電切系統(tǒng)常見(jiàn)的故障

(4) Common faults in electrocoagulation and electrocautery systems

(a)設(shè)備配件損壞。如腳踏開(kāi)關(guān),負(fù)極板導(dǎo)線(xiàn),連接導(dǎo)線(xiàn)等常用的易損配件。

(a) Damaged equipment accessories. Commonly used vulnerable accessories such as foot switches, negative electrode wires, and connecting wires.

(b)電切、電凝,雙極故障。在設(shè)備內(nèi)部,這些部分都由不同的電路模塊組成。我們可以根據(jù)其表現(xiàn)的故障現(xiàn)象對(duì)相應(yīng)的模塊電路進(jìn)行維修或更換。

(b) Electric cutting, electrocoagulation, bipolar fault. Internally, these parts are composed of different circuit modules. We can repair or replace the corresponding module circuit based on the symptoms of the malfunction.

本文來(lái)自:內(nèi)窺鏡設(shè)備維修更多的內(nèi)容請(qǐng)點(diǎn)擊:http://wufangbu88888.cn我們將會(huì)為您提問(wèn)的問(wèn)題提供一個(gè)滿(mǎn)意的服務(wù),歡迎您的來(lái)電!

This article is from: Endoscopic equipment maintenance. For more information, please click on: http://wufangbu88888.cn We will provide satisfactory service for the questions you ask. Welcome to call us!

(1)光源系統(tǒng)常見(jiàn)的故障

(1) Common faults in light source systems

(a)光線(xiàn)發(fā)暗。排除了是光柵遮擋的原因外,基本可以肯定是燈泡接近使用壽命,更換新燈泡則可。氙光燈泡的使用壽命一般設(shè)計(jì)在500小時(shí),如果光源系統(tǒng)有壽命計(jì)時(shí)裝置,這個(gè)問(wèn)題就很容易。

(a) The light is dim. Excluding the possibility of grating obstruction, it is generally certain that the bulb is nearing its lifespan, and replacing it with a new bulb is sufficient. The service life of xenon light bulbs is generally designed to be 500 hours. If the light source system has a lifespan timer device, this problem can be easily solved.

(b)開(kāi)機(jī)燈泡不亮。如果排除了燒掉燈泡這個(gè)原因的話(huà),那么問(wèn)題常常就出在主機(jī)高壓點(diǎn)火電路部分,因?yàn)殚_(kāi)機(jī)的時(shí)候機(jī)器會(huì)產(chǎn)生上萬(wàn)伏特的脈沖高壓,元件會(huì)老化得特別快。注意此時(shí)不能用萬(wàn)用表去檢查脈沖高壓。

(b) The light bulb does not light up when turned on. If the reason for burning out the light bulb is ruled out, the problem often lies in the high-voltage ignition circuit of the host, because when the machine is turned on, it will generate tens of thousands of volts of pulse high voltage, and the components will age very quickly. Be careful not to use a multimeter to check for pulse high voltage at this time.

(2)人工氣腹系統(tǒng)常見(jiàn)的故障

(2) Common malfunctions of artificial pneumoperitoneum systems

(a)壓力顯示不正確。這多是氣體壓力傳感器損壞,如果機(jī)器設(shè)計(jì)有一組多個(gè)壓差壓力傳感器同時(shí)工作的,一般一個(gè)老化或者污染無(wú)法工作的時(shí)候,我們要把這組傳感器全部更換,否則只換其中一個(gè),其他的同組的傳感器壽命也接近臨界了。

(a) The pressure display is incorrect. This is mostly due to damage to gas pressure sensors. If the machine is designed with a group of multiple differential pressure sensors working simultaneously, usually when one sensor is aging or contaminated and cannot work, we need to replace all of this group of sensors. Otherwise, only one of them will be replaced, and the lifespan of other sensors in the same group will also be close to the critical point.

(b)流量不準(zhǔn)確。氣體流量傳感器損壞更換即可。有時(shí)候是通過(guò)壓差來(lái)計(jì)算流量的,這時(shí)候我們就要更換新的壓差傳感器。

20200426035244690

(b) Inaccurate traffic. Replace the damaged gas flow sensor. Sometimes the flow rate is calculated by pressure difference, and at this time we need to replace it with a new pressure difference sensor.

(c)設(shè)備漏氣。很多情況是器械密封圈老化而引起的。但也有設(shè)備機(jī)械減壓閥漏氣和電磁閥損壞的原因,這個(gè)比較容易判斷出并對(duì)其進(jìn)行修復(fù)。

(c) Equipment leaks air. Many cases are caused by aging of instrument sealing rings. But there are also reasons for mechanical pressure reducing valve leakage and electromagnetic valve damage, which are relatively easy to identify and repair.

(3)液體加壓系統(tǒng)常見(jiàn)的故障

(3) Common faults in liquid pressurization systems

(a)蠕動(dòng)泵損壞。

(a) The peristaltic pump is damaged.

(b)液體壓力傳感器損壞。其原理基本與氣體壓力傳感器相同。

(b) The liquid pressure sensor is damaged. Its principle is basically the same as gas pressure sensors.

(c)設(shè)備內(nèi)部電路損壞。

(c) The internal circuit of the equipment is damaged.

(4)電凝電切系統(tǒng)常見(jiàn)的故障

(4) Common faults in electrocoagulation and electrocautery systems

(a)設(shè)備配件損壞。如腳踏開(kāi)關(guān),負(fù)極板導(dǎo)線(xiàn),連接導(dǎo)線(xiàn)等常用的易損配件。

(a) Damaged equipment accessories. Commonly used vulnerable accessories such as foot switches, negative electrode wires, and connecting wires.

(b)電切、電凝,雙極故障。在設(shè)備內(nèi)部,這些部分都由不同的電路模塊組成。我們可以根據(jù)其表現(xiàn)的故障現(xiàn)象對(duì)相應(yīng)的模塊電路進(jìn)行維修或更換。

(b) Electric cutting, electrocoagulation, bipolar fault. Internally, these parts are composed of different circuit modules. We can repair or replace the corresponding module circuit based on the symptoms of the malfunction.

本文來(lái)自:內(nèi)窺鏡設(shè)備維修更多的內(nèi)容請(qǐng)點(diǎn)擊:http://wufangbu88888.cn我們將會(huì)為您提問(wèn)的問(wèn)題提供一個(gè)滿(mǎn)意的服務(wù),歡迎您的來(lái)電!

This article is from: Endoscopic equipment maintenance. For more information, please click on: http://wufangbu88888.cn We will provide satisfactory service for the questions you ask. Welcome to call us!

維修項(xiàng)目
聯(lián)系我們
  • 公司公司:匠仁醫(yī)療設(shè)備有限公司
  • 匠仁醫(yī)療山東濟(jì)南維修部
  • 聯(lián)系人:樊經(jīng)理
  • 電話(huà):13153199508
  • 地址:山東省濟(jì)南市槐蔭區(qū)美里東路3000號(hào)德邁國(guó)際信息產(chǎn)業(yè)園6號(hào)樓101-2室
  • 匠仁醫(yī)療湖南長(zhǎng)沙維修部
  • 聯(lián)系人:李經(jīng)理
  • 電話(huà):13873135765
  • 地址:湖南省長(zhǎng)沙市雨花區(qū)勞動(dòng)?xùn)|路820號(hào)恒大綠洲14棟2409室
新聞資訊

樊經(jīng)理:13153199508     李經(jīng)理:13873135765

公司地址:山東省濟(jì)南市槐蔭區(qū)美里東路3000號(hào)德邁國(guó)際信息產(chǎn)業(yè)園6號(hào)樓101-2室  湖南省長(zhǎng)沙市雨花區(qū)勞動(dòng)?xùn)|路820號(hào)恒大綠洲14棟2409室

     

魯公網(wǎng)安備 37010402441281號(hào)