(1)腹腔鏡設(shè)備的清洗:術(shù)后應(yīng)立即拆卸腹腔鏡設(shè)備的可拆卸部分,并對其進(jìn)行清洗.先用酶劑浸泡,再用流動水沖洗.對帶管腔的器械,每次使用后應(yīng)使用通條刷反復(fù)通刷,再用流水或高壓水槍反復(fù)沖洗.
(1) Cleaning of laparoscopic equipment: the detachable part of laparoscopic equipment shall be removed immediately after operation and cleaned. It shall be soaked with enzyme and then washed with flowing water. For instruments with lumen, it shall be brushed repeatedly with a full brush after each use, and then washed repeatedly with running water or high-pressure water gun
對穿刺器械使用棉簽清洗閥門內(nèi)腔、彈簧、密封圈及穿刺錐的側(cè)孔,吹干后及時(shí)將閥門還原,以避免閥門內(nèi)的小部件遺失.清洗抓持器械及剪刀時(shí)應(yīng)打開旋轉(zhuǎn)軸輕輕刷洗.若軸節(jié)仍有頑固碎屑則采用超聲清洗器清洗.攝像頭、腹腔鏡視角用軟布擦干晾干,切忌用粗糙的布巾擦拭以免劃傷鏡頭,嚴(yán)禁使用超聲波清洗機(jī)對內(nèi)鏡進(jìn)行清洗.
Use a cotton swab to clean the inner cavity of the valve, the spring, the sealing ring and the side hole of the puncture cone. After drying, restore the valve in time to avoid the loss of small parts in the valve. When cleaning the grasping instrument and scissors, open the rotating shaft and brush gently. If there are still stubborn debris on the coupling, clean it with an ultrasonic cleaner. Wipe the camera and laparoscope with a soft cloth to dry, Never wipe with a rough cloth towel to avoid scratching the lens. It is strictly prohibited to use an ultrasonic cleaner to clean the endoscope
(2)腹腔鏡設(shè)備的:針對不同的器械可分別采用高溫高壓法、低溫法等進(jìn)行.攝像頭、電纜和電極不可使用液浸泡法或高溫高壓法,否則對器械會造成嚴(yán)重的損傷,只能采用氣體熏蒸法或等離子低溫法.
(2) Disinfection of laparoscopic equipment: different instruments can be disinfected by high-temperature and high-pressure sterilization and low-temperature sterilization respectively. Cameras, cables and electrodes cannot be disinfected by disinfectant immersion or high-temperature and high-pressure sterilization, otherwise serious damage will be caused to the instruments. Only gas fumigation or plasma low-temperature sterilization can be used